Polypropylene Case Pressure Gauge, Gujarat, India

response@pressuregaugeindia.com | +91 9426395647

Polypropylene Case Pressure Gauge

Polypropylene Case Pressure Gauge (4-½’’)

Bourdon type

Polypropylene Case Pressure Gauge Send Enquiry

Special Features

  • Polypropylene case with SS measuring system
  • Wetted parts AISI 316L stainless steel
  • Accuracy +0.5% F.S.
  • Solid front design features baffle wall interposed between the sensing system, the window and a pressure relieving back for increased safety
  • Standard followed ASME - B 40.100

Application

  • Suitable for high pressure gas
  • Corrosive / hazardous environment that will not obstruct the pressure system.

Specifications

Standard Version : 41/4”

Accuracy±1% F. S. (As per ASME B40.100)
Ambient temperature-20°C to +65°C
Process temperatureMax 150°C
Operating pressure range75% of Scale Value
Over pressure limit< 100 bar : 125% of Max. Scale Value

> 100 to < 600 bar : 115% of Max. Scale Value

> 600 to < 1600 bar: 110% of Max. Scale Value

CaseFibre Glass Reinforced Polypropylene
BourdonAISI 316LSS
SocketAISI 316LSS
MovementAISI 304 SS
JointsTig Argon Arc Welding

ProtectionIP 68
DialAluminium, black graduation on white background
PointerAluminium, black Coloured, Micrometer Zero Adjustable
WindowToughened Glass
Blow off DiscStainless Steel
GasketNeoprene

Glycerine Filled Version (Option LFG)

Accuracy: ±1.0% of F.S.
Ambient TemperatureMaximum 65°C
Process TemperatureMaximum 65°C
Dampening LiquidsGlycerine 99.7%

(Others available as Option)
Other FeaturesRefer Specification

of Standard Version
ProtectionIP 68

Dry But Fillable Version (Option DFG)

Fillable Dampening LiquidGlycerine 99.7%
Ambient TemperatureMaximum 65°C
Process TemperatureMaximum 65°C
Other FeaturesRefer Specification of Standard Version

Temperature effect:

The variation of indication caused by effects of temperature is to be calculated by below formula; which is to be added in the specified accuracy while measurementFormula : ±0.04 x (t2 -11) % of F. S. where t1 = reference temperature (+20°C) and t2 = ambient temperature in °C.

Download Brochure

Accessories

Bourdon Type

Diaphragm Sealed & Gauge

Differential Gauge

Digital Pressure and Temperature

Electrical Contact

Low Pressure Capsule & LPD & SCH Diaphragm

Manifold

Megnahellic DPG

Needle Valve

RTD

Special Gauges

Temperature Gauges

Thermowell